User guide. Read online or download PDF. Eppendorf 5424 R Centrifuge User Manual. Eppendorf Equipment Manuals Directory ManualsDir.com - online owner manuals library Search. If you are searched for the ebook Eppendorf centrifuge 5415r service manual in pdf form, then you have come on to faithful website. We presented the full variation of this ebook in DjVu, PDF, doc, ePub, txt formats. You can read Eppendorf centrifuge 5415r service manual online or load. Centrifuge 5425 is the new laboratory standard. With its 24–place capacity and speed up to 21,300 × g, it is perfectly equipped for all modern molecular biology applications in Eppendorf and PCR tubes.
31
7 Troubleshooting
Centrifuge 5424 / 5424 R — Operating manual
7.2
Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.
Symptom / message
Cause
Remedy
LID ERROR
Centrifuge lid cannot be locked.
Try to close the centrifuge lid again.
LID ERROR
Centrifuge lid cannot be released.
1. Switch the device off and back on.
2. Press the open key.
If the error occurs again:
1. Switch off the centrifuge.
2. Activate the emergency lid release
(see Emergency release on p. 32).
LID ERROR
Centrifuge lid must not be released during a
run.
Wait for the rotor to stop.
LID LIFT
(only 5424 R)
Centrifuge lid has not been opened wide
enough.
Open the centrifuge lid wider by hand.
INT
Mains power failure during a run.
Check the mains connection.
NO RPM
Error in speed measuring system.
Leave the device switched on until the
error message disappears (10 s or 6 min).
Err 6
Drive error.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on again
after > 20 s.
Err 6
Eppendorf 5424 R Service Manual
Drive overheated.
Allow the drive to cool down for at least 15
min.
Err 7
Major deviation in the speed control.
1. Wait for the rotor to stop.
2. Tighten the rotor.
Service Manual Tv
Err 8
Drive error.
1. Wait for the rotor to stop.
2. Repeat the run.
Err 9 to 17
Electronics error.
Switch the centrifuge off and back on
again after > 20 s.
Err 18
(only 5424 R)
Too high temperature variation from nominal
value in the rotor chamber.
Check the settings.
Check unhindered air circulation through
the vents.
Thaw ice or switch off device and allow it
to cool down.
Err 19
(only 5424 R)
Cooling circuit is overheated.
Eppendorf Centrifuge 5702 User Manual
Vic 2 ultimate mod download. Aug 17, 2014 Here is the download link for the Victoria Ultimate mod: Latest version is 0.43. Apr 12, 2016 Ultimate is an ambitious mod for Paradox Interactive's outstanding grand strategy game Victoria II, it has an expansive feature list with multiple new start dates, from 1604 to 2023.
Check unhindered air circulation through
the vents and allow device to cool down.
Err 20
(only 5424 R)
Temperature sensor in rotor chamber faulty.
Switch the centrifuge off and back on
again after > 20 s.
Err 21
(only 5424 R)
Temperature sensor at capacitor faulty.
Switch the centrifuge off and back on
again after > 20 s.
Eppendorf 5424 R Service Manual
Err 24
(only 5424 R)
Cooling unit fault, e.g., overheating.
Allow the centrifuge to cool down and
repeat the cycle.
EN
O
perating manual